注目情報はこちら >>
72782
7-AAD/CFSE Cell-Mediated Cytotoxicity Assay Kit
Cellular Assay Kits
Assay Kit

7-AAD/CFSE Cell-Mediated Cytotoxicity Assay Kit #72782

Citations (0)
Flow cytometric analysis of target K562 cells (top row) or SUP-B15 cells (bottom row) labeled with CFSE, incubated with (right column) or without (left column) effector NK-92 cells at a 12:1 effector:target cell ratio for 4 h, and stained with 7-AAD for 15 min. Cells that stain positive for both 7-AAD and CFSE represent dying target cells and are located in Q2 of the dot plot.
To Purchase # 72782
Cat. # Size Qty. Price Inventory
72782S
1 Kit  (100 assays)

Product Includes Quantity (with Count)
7-AAD Viability Dye 2 x 50 µl
CFSE Stock Solution 1 x 100 µl
Cell-Based Assay Buffer Tablet 1 x 4 ea

Protocol

PRINT

View >Collapse >

#72782 7-AAD/CFSE Cell-Mediated Cytotoxicity Assay Kit

A. Solutions and Reagents

Supplied Reagents:

  1. Cell-Based Assay Buffer: Dissolve each Cell-Based Assay Buffer Tablet (#27868) in 100 ml of reverse osmosis deionized (RODI) water. Mix well to ensure that the tablet dissolves completely. The diluted buffer is stable at room temperature for one year.
  2. 7-AAD Viability Dye: Add 10 μl of 7-AAD Viability Dye (1,000X) (#89587) to 10 ml of Cell-Based Assay Buffer and mix well.
  3. CFSE Staining Solution: First, prepare a 0.1% BSA/Assay Buffer by adding 10 mg BSA to 10 ml of Cell-Based Assay Buffer. Then make a 1:500 dilution of CFSE Stock Solution (#16650) in 0.1% BSA/Assay Buffer and mix well. You will need 1 ml of CFSE Staining Solution per 107 cells.

NOTE: The recommended protocol labels cells with CFSE brightly enough to distinguish labeled and unlabeled cells, but dimly enough not to bleed into other channels. However, should brighter staining be desired, BSA can be omitted from the CFSE Staining Solution preparation.

Additional Reagents (Not Supplied):

  1. Bovine Serum Albumin (BSA) (#9998)

B. Labeling of Target Cells

NOTE: To properly analyze the data, the following control target cell groups are needed to set up the flow cytometer and compensation:

  • Unstained target cells
  • Single-stained target cells for each label or stain

  1. Obtain target cells for your cytotoxicity assay. Optimal conditions and incubation times for this assay should be determined on an individual basis.
  2. Centrifuge the cells at 400 x g for five minutes. Aspirate the supernatant and flick the tube well to break up the pellet.
  3. Quickly resuspend cell pellet in CFSE Staining Solution (prepared as described in Solutions and Reagents) at a concentration of 107 cells/ml. A uniform suspension should be reached as quickly as possible, as CFSE is taken up almost immediately and local variations in CFSE concentrations can affect staining uniformity. Control target cells (target cells without CFSE) should be resuspended in 0.1% BSA/Assay Buffer.
  4. Incubate the cells in the CFSE Staining Solution for 15 minutes at 37°C.
  5. Add at least 10 volumes of culture medium containing FBS. Centrifuge the target cells at 400 x g for five minutes.
  6. Aspirate the supernatant.
  7. Resuspend the target cells in 10 ml of culture medium.
  8. Centrifuge the target cells at 400 x g for five minutes.
  9. Aspirate the supernatant.
  10. Resuspend the target cells in culture medium at a concentration of 105 cells/ml.
  11. Incubate the cells at 37°C for 30 minutes or longer (but not long enough for the cells to proliferate) in a CO2 incubator.

C. Assay Procedure

  1. Collect effector cells into tubes. Centrifuge the cells at 400 x g for five minutes to pellet.
  2. Resuspend the cells in culture medium at a concentration of 5 x 106 cells/ml.
  3. Add effector cells to the CFSE-labeled target cell suspension at a predetermined effector/target cell ratio. Some examples are shown in the table below:

    Effector:Target Ratio Effector Cell Suspension Target Cell Suspension Target Cell Medium Final Volume
    0 0 ml 1.5 ml 0 ml 1.5 ml
    6.25:1 0.125 ml 1 ml 0.375 ml 1.5 ml
    12.5:1 0.25 ml 1 ml 0.25 ml 1.5 ml
    25:1 0.5 ml 1 ml 0 ml 1.5 ml
  4. Incubate the cell mixture for four hours or for a period of time according to your optimal protocol, allowing enough time for cytolytic activity to progress.
  5. To stain in a 96 well v-bottom plate as described here, transfer cells into the plate and centrifuge at 400 x g for five minutes. (For staining in tubes, scale volumes up 5-fold).
  6. Aspirate the supernatant.

    NOTE: If additional surface markers are to be assayed, staining can be inserted at this point in the protocol.

  7. Resuspend the cells in 200 μl of 7-AAD Viability Dye (prepared as described in Solutions and Reagents) and mix well. The control target cells (target cells without CFSE or 7-AAD viability dye or target cells with CFSE staining only) should be resuspended in 200 μl of Assay Buffer.
  8. Incubate the cells for 15 minutes in the dark at 4°C.
  9. The cells are now ready for analysis with a flow cytometer and should be analyzed immediately. Gate on CFSE+ target cells (ex/em 488/525), and then visualize the live/dead cell percentages by 7-AAD exclusion (ex/em 488/647).

D. Troubleshooting


Problem Possible Causes Recommended Solutions
Low signal of CFSE
  1. Cells are not healthy
  2. Cells were not well labeled by CFSE
  1. Use only healthy cells
  2. Perform a titration of CFSE to get an optimal concentration of CFSE staining
No difference in cytotoxicity among different effector cell to target cell ratios
  1. Target cells are not healthy
  2. Effector cells do not cause cytotoxicity
  1. Use only healthy target cells
  2. Use effector cells that cause cytotoxicity

Posted February 2020

Protocol Id: 1964

Product Description

The 7-AAD/CFSE Cell-Mediated Cytotoxicity Assay Kit employs CFSE (carboxyfluorescein succinimidyl ester) to label target cells within the mixed cell population and 7-AAD (7-aminoactinomycin D) to label dead cells. This kit provides an improvement over the traditional chromium-51 (51 Cr) release assay to assess cell-mediated cytotoxicity. CFSE labeling is more sensitive, does not employ radioisotopes, and cytolysis can be assessed at the single-cell level using flow cytometry.

Background

Cytolytic assays are routinely used to interrogate immune effector cell function. In order to interrogate immune effector cell cytolytic activity in a heterogeneous cell population of effector and target cells, it is imperative to be able to discriminate between effector and target cell populations with distinct phenotypes. As such, this kit is designed to label live target cells with the membrane permeable fluorescent dye CFSE prior to coculture with unlabelled effector cells, thus permitting a clear separation between live effector and target cells. After incubation of live effector and target cells, the DNA intercalating dye 7-AAD is added to label dying target cells with compromised plasma membranes. By using dyes with distinct spectral properties, a clear separation between four cell populations can be obtained using flow cytometry: live target cells, dead target cells, live effector cells, and dead effector cells (1).

Limited Uses

Except as otherwise expressly agreed in a writing signed by a legally authorized representative of CST, the following terms apply to Products provided by CST, its affiliates or its distributors. Any Customer's terms and conditions that are in addition to, or different from, those contained herein, unless separately accepted in writing by a legally authorized representative of CST, are rejected and are of no force or effect.

Products are labeled with For Research Use Only or a similar labeling statement and have not been approved, cleared, or licensed by the FDA or other regulatory foreign or domestic entity, for any purpose. Customer shall not use any Product for any diagnostic or therapeutic purpose, or otherwise in any manner that conflicts with its labeling statement. Products sold or licensed by CST are provided for Customer as the end-user and solely for research and development uses. Any use of Product for diagnostic, prophylactic or therapeutic purposes, or any purchase of Product for resale (alone or as a component) or other commercial purpose, requires a separate license from CST. Customer shall (a) not sell, license, loan, donate or otherwise transfer or make available any Product to any third party, whether alone or in combination with other materials, or use the Products to manufacture any commercial products, (b) not copy, modify, reverse engineer, decompile, disassemble or otherwise attempt to discover the underlying structure or technology of the Products, or use the Products for the purpose of developing any products or services that would compete with CST products or services, (c) not alter or remove from the Products any trademarks, trade names, logos, patent or copyright notices or markings, (d) use the Products solely in accordance with CST Product Terms of Sale and any applicable documentation, and (e) comply with any license, terms of service or similar agreement with respect to any third party products or services used by Customer in connection with the Products.

For Research Use Only. Not for Use in Diagnostic Procedures.
Cell Signaling Technology is a trademark of Cell Signaling Technology, Inc.
All other trademarks are the property of their respective owners. Visit our Trademark Information page.